ISO 5347-17:1993
ISO 5347-17:1993
Methods for the calibration of vibration and shock pick-ups — Part 17: Testing of fixed temperature sensitivity
Edition 1
1993-12
Withdrawn
ISO 5347-17:1993
11363
Withdrawn (Edition 1, 1993)

Abstract

Lays down detailed specifications for the instrumentation and procedure to be used for fixed temperature sensitivity testing. Applies to rectilinear pick-ups. Is applicable for the following parameters: frequency range: 20 Hz to 1250 Hz; dynamic range: 0,1 µm to 10 mm (frequency-dependent), 1 mm/s to 1 m/s (frequency-dependent), 10 m/s^2 to 1000 m/s^2 (frequency-dependent); temperature range: - 45 °C to + 800 °C. The uncertainty applicable is ± 10 % of reading.

General information

  •  : Withdrawn
     : 1993-12
    : Withdrawal of International Standard [95.99]
  •  : 1
     : 2
  • ISO/TC 108
    17.160 